Table 5 Percentage of facilities that implemented four measures and percentage of facilities where equipment was damaged in three prefectures

From: Survey of dialysis therapy during the Great East Japan Earthquake Disaster and recommendations for dialysis therapy preparation in case of future disasters

  Fastening dialysis fluid delivery systems and RO systems to floor (%) Using flexible tubes for connection (%) Dialysis monitors (%) Dialysis beds (%) Equipment damage (%)
Miyagi 92.5 92.5 87.8 85.2 14.3
Fukushima 73.2 58.9 90.9 86.2 26.4
Ibaraki 43.8 50.0 85.8 93.3 11.4
  1. Additional survey of 298 facilities