Skip to main content

Table 7 Endotoxin levels and bacterial numbers immediately before, after, and approximately 1 year after ETRF installation in Cambodia

From: Present status of renal replacement therapy at 2015 in Asian countries (Myanmar, Vietnam, Thailand, China, and Japan)

  Endotoxin (EU/mL) Bacterial colony (CFU/mL)
Before ETRF installation
 RO water 0.120 180
 Dialysis fluid 0.650 >300
After ETRF installation
 Dialysis fluid <0.001 <0.1
Approximately 1 year after ETRF installation
 RO water 0.733 >300
 Dialysis fluid <0.001 <0.1