Table 1 Technical data on BG and NF

From: A new polymethylmetacrylate membrane improves the membrane adhesion of blood components and clinical efficacy

  BG NF
Ultrafiltration ratea [mL/h mmHg] 38 35
Clearanceb [mL/min] Urea 191 193
Creatinine 174 180
Phosphate 164 172
Vitamin B12 114 126
Negative chargec [meq/g] 110 80
  1. aUltrafiltration rate was measured with bovine blood (hematocrit, 30 ± 3%; total protein, 6.0 ± 0.5 g/dl; blood flow rate, 200 ± 4 mL/min; transmembrane pressure, 13.3 ± 1.3 kPa; temp., 37 ± 1 °C
  2. bClearance was measured with aqueous solution. Blood flow rate, 200 ± 4 mL/min; dialyzate flow rate, 500 ± 10 mL/min, filtration rate, 10 ± 2 mL/min; temp., 37 ± 1 °C
  3. cNegative charge on the membranes was determined from the evaluation of the triiodide complex. It was created by the reaction of iodine and iodide ions after dried hollow fiber membrane immersed in a solution of 5% potassium iodide in methanol for 24 h