Fig. 2 | Renal Replacement Therapy

Fig. 2

From: Effect of hollow fiber diameter and membrane surface area of polymethyl methacrylate membrane on filter lifetime

Fig. 2

TMP A and pressure drop B when the pressure was stable in all the experiments. The TMP was significantly lower for the CH-1.8W filter than for the CH-1.0W and CH-1.0N filters. The pressure drop was significantly lower for the CH-1.8W filter than for the CH-1.0W and CH-1.0N filters, and that for the CH-1.0W filter was lower than that for the CH-1.0N filter