Fig. 3 | Renal Replacement Therapy

Fig. 3

From: Effect of hollow fiber diameter and membrane surface area of polymethyl methacrylate membrane on filter lifetime

Fig. 3

Times at which the TMP reached 200 mmHg A, the dP/dt of the TMP reached 15 mmHg/h B, the pressure drop across the hemofilter reached 200 mmHg C, and the dP/dt of the pressure drop reached 15 mmHg/h D. The times at which the TMP reached 200 mmHg, the dP/dt of the TMP reached 15 mmHg/h, and the pressure drop reached 200 mmHg were significantly longer for the CH-1.8W filter than for the CH-1.0N and CH-1.0W filters. The time at which the dP/dt of the pressure drop reached 15 mmHg/h was significantly longer for the CH-1.8W filter than for the CH-1.0N filter