Table 2 Effects of flow stoppage time and the presence of metal corrosion on number of detection times from the metal parts

From: Effects of metal corrosion in the pump of a dialysis machine on the sterility of the terminal dialysate by spike-and-recovery testing with bacteria

Metal corrosion

Flow stoppage time (h)

Detected times

Not detected times

p-Value (12 h versus 18 h)

Absent

12

1

11

 < 0.01

18

10

2

Present

12

3

9

 < 0.01

18

12

0

  1. Dialysate spiked with P. aeruginosa (10 CFU/mL) was introduced to the experimental circuit (Fig. 1), and stopped flow for indicated times. After the cleaning and disinfection as shown in Fig. 4, the dialysis pump was disassembled, and bacteria on 12 metal parts were released by ultrasonic exposure. Released live bacteria were detected. **p < 0.01 (Holm’s test)