Fig. 3 | Renal Replacement Therapy

Fig. 3

From: Differences in chamber structure contribute to the incidence of venous air trap chamber coagulation during AN69ST-CHF: clinical and in vitro evaluation

Fig. 3

Experiment 1: investigation of the dye disappearance time in a venous air trap chamber. The dye disappearance time in the venous air trap chamber was significantly shorter when using Prismaflex (17.5 s ± 0.7) than when using ACH-Σ (51.2 s ± 0.7; p < 0.05). In the Prismaflex group, fluid throughout the entire chamber was mixed and no retention was observed, but in the ACH-Σ group, retention in the absence of fluid flow was observed in the upper layer of the accumulated liquid