Table 4 SiO2 concentration in various portions of CDDS and in commercial sterile solutions

From: Particulates formed in dialysis fluid: effect of silica in tap water

 

SiO2 (mg/L)

Final ETRF use time (month or months)

1

2

3

4

5

6

CDDS

 Tap water

6.1

5.8

6.1

7.7

7.2

6.7

 RO permeate

0.5

0.4

0.5

0.7

0.7

0.7

 Dialysis water (substitution fluid sampling point)

0.5

0.4

0.5

0.7

0.7

0.7

Commercial sterile acidic substitution fluid

0.1

Commercial sterile bicarbonate substitution fluid

0.1

5% glucose injection

Below the lower limit of quantification

  1. SiO2 in tap water used for CDDS, RO permeate, dialysis water, commercial sterile acidic substitution fluid and commercial sterile bicarbonate substitution fluid, and 5% glucose injection. CDDS central dialysis fluid delivery system, ETRF endotoxin-retentive filters, SiO2 silica