Particulates formed in dialysis fluid: effect of silica in tap water

Abstract

Background

A fluid commonly used for dialysis in Japan is prepared from two dialysate concentrates and dialysis water. Silicon (Si) from tap water is reported in dialysis water as suspended insoluble particulate matters. We aimed to examine the source of insoluble particulate matters in the online prepared substitution fluid and to examine effects of the silica (SiO2) of tap water on the SiO2 of dialysis water.

Method

Based on the number of particles and particle size distribution in online prepared substitution fluid, we evaluated particle diameters and constituent elements of particles in online prepared substitution fluid. To identify sources of insoluble particulate matters in the online prepared substitution fluid, we measured particle diameters and constituent elements of particles adhering to the surface of hollow fiber of the final endotoxin-retentive filter (ETRF) and to the inner surface of the housing of the final ETRF and measured SiO2 in tap water, dialysis water, and commercial sterile substitution fluid.

Results

Si-containing particles accounted for 83% of 30 particles detected in the final ETRF (the hollow fibers and the housing) and the online prepared substitution fluid sampled from the final ETRF. The large variability in the SiO2 in tap water depending on the geographic regions. The SiO2 in dialysis water was 0.4–0.7 mg/L. Si was not detected in particles in the commercial sterilized substitution fluid, and the commercial sterile substitution fluids contained SiO2 at 0.1 mg/L.

Conclusion

We presume that insoluble particulates in online prepared substitution fluid were calcium and magnesium precipitated from dialysate and Si precipitated from dialysis water. Moreover, variation in the SiO2 in tap water is a factor affecting the SiO2 level in dialysis water. If the formation of silicates in the online prepared substitution fluid is considered and the target SiO2 in dialysis water is to be set ≤ 0.1 mg/L, SiO2 in tap water available in hospitals and dialysis centers should be known before specifications of instruments for dialysis fluid preparation are determined.

Background

A fluid commonly used for dialysis in Japan is prepared from two dialysate concentrates (acid, A; bicarbonate, B) and dialysis water. The dilution ratio is dialysate concentrate A/dialysate concentrate B/dialysis water = 1:1.26:32.74. Moreover, the dialysate concentrates A and B are prepared by dissolving the powdery dialysate concentrates in dialysis water. Therefore, the quality of dialysis fluid largely depends on the quality of dialysis water [1]. While the powder dialysate concentrates are manufactured in accordance with the Japanese Pharmacopoeia (JP) [2], dialysis water is primarily manufactured from tap water; therefore, the Japanese Society for Dialysis Therapy (JSDT) has proposed two sets of standards for dialysis water, standard established for biological contamination and water quality standard for chemical contaminants [3]. However, silicon (Si) from tap water is reported in dialysis water as suspended insoluble particulate matters [4]. Furthermore, the JSDT standards do not reference insoluble particulate matters in dialysis water or dialysis fluid.

In online hemodiafiltration (HDF) therapy, 10–100 L of online prepared substitution fluid per dialysis session is infused into the blood of a patient undergoing extracorporeal circulation [5]. An online prepared substitution fluid is produced by filtration, through an endotoxin-retentive filter (ETRF), of a dialysis fluid [6]. Based on the ETRF specifications, insoluble particulate matters > 0.1 µm are unlikely through ETRF; however, a study showed the presence of insoluble particulate matters in dialysis fluids filtered with an ETRF [4]. With the allowance that the injection volume of online prepared substitution fluid is used in a larger quantity than commercial sterile substitution fluid, even the trivial amount of insoluble particulate matters might lead to eventual health hazard.

In the present study, we first measured the number of particles and particle size distribution in online prepared substitution fluid prepared using ETRF on a bedside dialysis console of the central dialysis fluid delivery system (CDDS) [7] that is cleaned using peracetic acid, commercial sterile substitution fluid, and 5% glucose injection. Referring to the JP and a special report by the American Society for Parenteral and Enteral Nutrition [8], we assessed the number of particles per 1 mL of the sample based on the following four categories of particle diameters: < 5 µm, ≥ 5 µm, ≥ 10 µm, and ≥ 25 µm. Based on the number of particles and particle size distribution, we evaluated the particles in online prepared substitution fluid using the highest number, the particles in commercial sterile substitution fluid, and 5% glucose injection using SEM–EDS that combines scanning electron microscopy (SEM) and energy-dispersive X-ray spectrometry (EDS). To identify sources of insoluble particulate matters in the online prepared substitution fluid, we measured particle diameters and constituent elements of particles adhering to the surface of hollow fiber of the final ETRF and to the inner surface of the housing of the final ETRF using SEM–EDS and measured silica (SiO2) in tap water, reverse osmosis (RO) water, dialysis water, commercial sterile acidic substitution fluid, commercial sterile bicarbonate substitution fluid, and 5% glucose injection with inductively coupled plasma optical emission spectroscopy (ICP-OES). In addition, since Matsumoto et al. have reported large differences in the Si level in tap water for preparing dialysis water depending on geographical regions [9], we determined SiO2 in tap water samples collected in various regions of Japan.

In this manner, we aimed to examine the source of insoluble particulate matters in the online prepared substitution fluid and to examine effects of the SiO2 of tap water on the SiO2 of dialysis water. These data are expected to establish specifications of instruments for preparing dialysis fluid in hospitals and dialysis centers to offer online prepared substitution fluid to provide the same level with the commercial sterile substitution fluid.

Methods

Central dialysis fluid delivery system

CDDS consists of a water treatment equipment (VCR-81NF-RUHS; Daicen Membrane Systems, Tokyo, Japan), a multipatient dialysis fluid supply unit (MDSU) (DAB-20NX; NIKKISO, Tokyo, Japan), an acid powder mixing device (AHI-502; DKK-TOA, Tokyo, Japan), a bicarbonate powder mixing device (BHI-502, DKK-TOA, Tokyo, Japan), and 12 bedside dialysis consoles (DCS-100NX; NIKKISO, Tokyo, Japan). The recovery rate of the reverse osmosis unit is approximately 75%, and the mean conductivity is 4.4 µS/cm. A bedside dialysis console is equipped with two ETRFs in series. The dialysis fluid piping has two systems—22 m and 15 m long—and each system has six bedside dialysis consoles connected. For the powder of dialysis fluid concentrate, Kindaly® 4E (Fuso Pharmaceutical Industries, Ltd., Osaka, Japan) is used. The CDDS disinfection method was post-dialysis rinsing for 30 min, followed by disinfection with sodium hypochlorite including a chelating agent (DIACLEAN; Kao, Tokyo, Japan; used after 140-fold dilution) for 40 min. After 40 min of disinfection, sodium hypochlorite including a chelating agent was retained overnight. Then, pre-dialysis rinsing was performed for 60 min. This disinfection process was performed three times a week. The method for acid cleaning of CDDS was post-dialysis rinsing for 30 min, followed by disinfection with sodium hypochlorite including a chelating agent (DIACLEAN; Kao, Tokyo, Japan; used after 140-fold dilution) for 40 min. After rinsing with dialysis water for 30 min, the system was cleaned with peracetic acid (SURGITECT; Kao, Tokyo, Japan; used after 100-fold dilution) for 40 min, and the peracetic acid solution was retained overnight. Then, pre-dialysis rinsing was performed for 60 min. This process was performed three times a week. Under these conditions, there were no CDDS issues because of carbonate.

Materials

We mixed the commercial sterile acidic substitution fluid and the commercial sterile bicarbonate substitution fluid used for offline HDF (SUBPACK-Bi®; NIPRO, Osaka, Japan) and heated the mixture for 20 min in a constant temperature water bath at 80 °C. The commercial sterile substitution fluid heated to 36 °C and 5% glucose injection (OTSUKA GLUCOSE INJECTION; Otsuka Pharmaceutical, Tokyo, Japan) were the target of the study.

Measuring the number of particles

The number of particles was measured under the following conditions according to the JP standard [2]. It was confirmed that the environment for the measurement was in accordance with the JP standard before counting particles. To remove foreign substances adhering to the beaker for sampling, water for Injection (OTSUKA DISTILLED WATER; Otsuka Pharmaceutical, Tokyo, Japan) was used for cleaning. Particles were counted using a light obscuration particle counter (KL-04; RION, Tokyo, Japan) under ventilation stopped. Thresholds for particle diameter were set at 1.3, 2.0, 5.0, 10.0, 20.0, 25.0, 40.0, 50.0, and 100.0 µm. For each measurement, 5 mL of sample was used. After four measurements, the first result was excluded, and the second to the fourth were averaged. We assessed the number of particles per 1 mL of the sample based on the following four categories: particle diameters of < 5 µm, ≥ 5 µm, ≥ 10 µm, and ≥ 25 µm.

We randomly chose one dialysis console out of 12 consoles equipped with the final ETRF (EF-02; NIKKISO, Tokyo, Japan) with 1–6 months of use. We passed dialysis fluid through the final ETRF on the selected bedside dialysis console for 15 min (dialysis fluid flow rate of 500 mL/min). After washing beakers with the sample liquid three times, we collected 300 mL of each online prepared substitution fluid, the commercial sterile substitution fluid, and the 5% glucose injection and left the samples for 30 min to avoid the impact of gas bubbles with a size of ≥ 2 µm before counting particles [10].

Measurement of particle diameters and constituent elements in particles

At the same time as sampling of liquid for particles counting, we collected 20 mL of each online prepared substitution fluid, the commercial sterile substitution fluid, and the 5% glucose injection in 50 mL centrifuge tubes (50-mL Super Clear Centrifuge Tubes 3191–335; Labcon North America, California, USA) after rinsing three times. Particles were captured with a membrane filter (0.4 μm pore size, FJ-VKF04; Koyo Sangyo, Tokyo, Japan) under the same conditions as counting of particles. To prevent precipitation of components onto the membrane filter, 20 mL of each liquid in the centrifuge tube and 50 mL of water for Injection were placed on the membrane filter while mixing, and particles from each liquid in the centrifuge tube were captured on the membrane filter. Our targets were particles in online prepared substitution fluid with the largest number. After suction, we coated the membrane filter with particles with osmium. We measured the particle diameter and elemental analysis of 10 particles on the coated membrane filter using SEM–EDS (accelerating voltage 8 kV) that combines SEM (Supra40; Carl Zeiss Microscopy GmbH, Jena, Germany) and EDS (XFlash 6|10; Bruker, Massachusetts, USA).

Observation of the final ETRF with SEM–EDS

After collecting the online prepared substitution fluid, we rinsed the CDDS with the dialysis water for 90 min to prevent precipitation of residual dialysis fluid components from the dialysis fluid piping. After rinsing, the final ETRF was removed from the dialysis console. This final ETRF was cut at 6 cm from the dialysis fluid inlet, and then it was air-dried until there was no remaining moisture. From six final ETRF that were treated in this manner, we selected the final ETRF that was directly connected to the sampling point for online prepared substitution fluid with the highest number of particles. We coated 10 central fibers of the selected final ETRF and 10 hollow fibers closest to the housing of the final ETRF with osmium. We analyzed the 2 cm of the osmium-coated hollow fibers on the dialysis fluid inlet side with SEM–EDS (accelerating voltage 8 kV). Referring to the result of particle numbers, we measured particle diameters and performed elemental analysis for 10 particles each. We also collected the housing of the selected final ETRF and coated it with osmium. We measured particle diameter and performed elemental analysis on the total of 10 particles adhering to the inner surface of the osmium-coated housing using SEM–EDS (accelerating voltage 8 kV) while referring to the result of particle numbers.

SiO 2 measurement

After the CDDS was rinsed with dialysis water for at least 90 min, water samples were collected in polyethylene bottles from the following three points: tap water used for the CDDS, the water treatment equipment (RO permeate), and the substitution fluid sampling point (dialysis water) (100 mL each) (location of collection, Bunkyo-ku, Tokyo; collection period, April–September 2017). To compare SiO2 in tap water, RO permeate, and dialysis water samples with SiO2 in commercial drugs, commercial sterile acidic substitution fluid, commercial sterile bicarbonate substitution fluid, and 5% glucose solution (100 mL each) were collected in polyethylene bottles. Furthermore, to compare tap water in various geographic regions of Japan, tap water samples were collected in polyethylene bottles at Kita-ku, Sapporo; Aoba-ku, Sendai; Bunkyo-ku, Tokyo; Nakamura-ku, Nagoya; Yodogawa-ku, Osaka; Naka-ku, Hiroshima; and Hakata-ku, Fukuoka (100 mL each) (collection period, the first week of April 2017). Polyethylene bottles cleaned with purified water were washed in advance with the solution to be collected three times. SiO2 in the tap water, RO permeate, dialysis water, commercial sterile acidic substitution fluid, commercial sterile bicarbonate substitution fluid, and 5% glucose solution collected as above were measured by ICP-OES (730-ES; Agilent Technologies, CA, USA) according to the Standard Methods for the Examination of Water of the Japan Water Works Association [11].

Results

Number of particles

Table 1 shows particle counts in the online prepared substitution fluid, commercial sterilized substitution fluid, and 5% glucose solution samples. Particles whose diameter ≥ 5 µm were detected in all samples. The largest number of particles was detected in the commercial sterilized substitution fluid. Among the online prepared substitution fluid samples, the largest number of particles was reported in the sample from the final ETRF tested after 3 months of use.

Table 1 Number of particles

Particle diameter and elemental analysis

Figure 1 shows SEM images and EDS mapping results for particles in the online prepared substitution fluid sample collected from the final ETRF after 3 months of use, the commercial sterilized substitution fluid sample, and the 5% glucose solution sample. Oxygen (O) and Si were detected in particles in the online prepared substitution fluid collected from the final ETRF after 3 months of use. Carbon (C) and O were detected in particles in the commercial sterilized substitution fluid and 5% glucose solution samples. Table 2 shows the particle diameter measurement and constituent element analysis results for 10 particles in the online prepared substitution fluid sampled from the final ETRF after 3 months of use, the commercial sterilized substitution fluid, and the 5% glucose solution. Of the particles in the online prepared substitution fluid sample collected from the final ETRF used for 3 months, three particles were composed of C, O, and calcium (Ca), three particles were composed of O and Si, and four particles were composed of O, magnesium (Mg), and Si. All ten particles in the commercial sterilized substitution fluid sample or the 5% glucose solution sample were composed of C and O.

Fig. 1
figure 1

SEM image and EDS mapping of particles in online prepared substitution fluid. SEM image and EDS mapping of particles in online prepared substitution fluid sampled from the final ETRF after 3 months of use, commercial sterile substitution fluid, and 5% glucose injection. In EDS mapping (elemental analysis), brown is C, pink is O, and turquois blue is Si. SEM, scanning electron microscopy; EDS, energy-dispersive X-ray spectrometry; ETRF, endotoxin-retentive filters

Table 2 Particle diameter and constituent elements analysis

Diameter and constituent elements of particles adhering to the final ETRF

Figure 2 shows SEM images and EDS mapping results for the surface of central fibers from the final ETRF used for 3 months, the surface of hollow fibers closest to the housing of the final ETRF, and the inner surface of the housing of the final ETRF. No particles were found on the surface of central fibers in the final ETRF. O, Mg, and Si were detected in the particles on the surface of the hollow fibers closest to the housing of the final ETRF. O and Si were detected in the particles on the inner surface of the housing of the final ETRF. Table 3 shows the particle diameter measurement and constituent element analysis results for ten particles on the surface of the hollow fibers closest to the housing of the final ETRF used for 3 months and ten particles on the inner surface of the housing of the final ETRF. All of the ten particles on the surface of the hollow fibers closest to the housing of the final ETRF used for 3 months were composed of O, Mg, and Si. Of the ten particles on the inner surface of the housing of the final ETRF, five particles were composed of O and Si, two particles were composed of C, O, and Ca, and three particles were composed of O, Mg, and Si.

Fig. 2
figure 2

SEM image and EDS mapping in the final ETRF. SEM image and EDS mapping of the central fiber surface, hollow fiber surface closest to the housing, and internal surface of the housing for the final ETRF with 3 months of use. In EDS mapping (elemental analysis), brown is C, pink is O, blue is Mg, turquois blue is Si. SEM, scanning electron microscopy; EDS, energy-dispersive X-ray spectrometry; ETRF, endotoxin-retentive filters

Table 3 Particle diameter and constituent of particles adhering to the final ETRF

SiO 2

Table 4 shows SiO2 in tap water used for CDDS, RO permeate, dialysis water, commercial sterile acidic and bicarbonate substitution fluids, and 5% glucose solution. The lower limit of quantification of SiO2 was 0.1 mg/L. The SiO2 in tap water and RO permeate was 5.8–7.7 and 0.4–0.7 mg/L, respectively. The SiO2 in dialysis water samples collected from the substitution fluid sampling point once a month until 6 months after the use of the final ETRF was started was 0.4–0.7 mg/L. Both the commercial sterile acidic and bicarbonate substitution fluids contained SiO2 at 0.1 mg/L. The SiO2 in the 5% glucose solution was below the lower limit of quantification. Figure 3 shows the SiO2 in tap water samples collected in various regions of Japan. The tap water samples from Sapporo, Sendai, Tokyo, and Fukuoka tended to contain higher levels of SiO2, and the tap water samples from Nagoya, Osaka, and Hiroshima tended to contain lower levels of SiO2.

Table 4 SiO2 concentration in various portions of CDDS and in commercial sterile solutions
Fig. 3
figure 3

SiO2 contents in tap water samples collected in various regions of Japan

Discussion

Formation mechanism of insoluble particulate matters in the final ETRF

Among the online prepared substitution fluid samples, the largest number of particles was detected in the online prepared substitution fluid sample collected from the final ETRF after 3 months of use. 1.7 particles with a diameter of 5 µm or larger were detected per 1 mL of the online prepared substitution fluid sampled from the final ETRF used for 3 months. Single-use filters, used for infusion therapy in critically ill patients, are effective in preventing particulates from entering the body and reduce the incidence of systemic inflammatory response syndrome (SIRS) [12]. However, ETRFs for preparing ultrapure dialysis fluid are reused in various countries [13].

Figure 4 shows a schematic of insoluble particulate matters adherence to the final ETRF and discharge into the online prepared substitution fluid. In the final ETRF, filtration was made from inside the hollow fibers to the outside. No particles on the surface of the central fibers in the final ETRF were observed. In the rinse, peracetic acid cleaning, and disinfection processes, the fluids from a central fiber in the final ETRF rinse, peracetic acid-clean, and disinfect the surface of other central hollow fibers, and insoluble particulate matters are possibly prevented from adhering to the surface of the central hollow fibers. Furthermore, no particles were observed in the central part of the final ETRF because, even if insoluble particulate matters are formed, they flow out from the center to the housing during each process.

Fig. 4
figure 4

Schematic diagram of insoluble particulate matters adhering to the final ETRF. Schematic diagram of insoluble particulate matters adhering to the final ETRF and discharging to online prepared substitution fluid. In constituent elements, C is carbon, O is oxygen, Mg is magnesium, Si is silicon, and Ca is calcium. ETRF stands for endotoxin-retentive filters

Particles composed of O, Mg, and Si were detected on the surface of the hollow fibers closest to the housing of the final ETRF. In the presence of ionic Mg, SiO2 precipitates as an Mg-containing silicate (magnesium silicate) even at low SiO2 concentrations [14]. Table 4 shows that the SiO2 in the dialysis water is 4–7 times higher than that in the commercial sterile acidic or bicarbonate substitution fluid, while 89–93% of SiO2 in tap water is removed by the water treatment equipment. The SiO2 in the 5% glucose solution was below the lower limit of quantification. These results suggest that magnesium silicate is prone to be formed in the online prepared substitution fluid. Moreover, only insoluble magnesium silicate [15] was attached to the hollow fibers closest to the housing of the final ETRF because these hollow fibers undergo less effective rinse, peracetic acid cleaning, and disinfection from the housing side compared with the central part of the final ETRF.

Particles composed of O and Si, C, O, and Ca, or O, Mg, and Si were detected on the inner surface of the housing of the final ETRF. Particles composed of O and Si are possibly amorphous SiO2, which is formed by independent precipitation of SiO2 in water [15]. Particles composed of C, O, and Ca are possibly CaCO3 because the formation of Ca-containing carbonate (CaCO3) overrides the formation of Ca-containing silicate [16]. The detection of only magnesium silicate on the surface of the hollow fibers closest to the housing of the final ETRF suggests that the inner surface of the housing undergoes less effective rinse, peracetic acid cleaning, and disinfection compared with the central part of the final ETRF and thus is prone to the formation of particles. Moreover, it is possible that insoluble particulate matters formed on the surface of the central fibers or the hollow fibers closest to the housing of the final ETRF were discharged during rinse, peracetic acid cleaning, and disinfection processes and attached to the inner surface of the housing.

The substitution fluid sampling point where online prepared substitution fluid was collected is directly connected to the final ETRF; therefore, entrance of particles of CaCO3, amorphous SiO2, and magnesium silicates to the body of a patient undergoing online HDF is unavoidable, as long as the online prepared substitution fluid is prepared using reused ETRF.

Particles in the commercial sterile substitution fluid and 5% glucose injection, and Si-containing particles in online prepared substitution fluid

The largest number of particles was detected in the commercial sterilized substitution fluid. Given a relatively low ionization degree of sodium hydrogen carbonate in commercial sterilized substitution fluids (0.88 at 36 °C), sodium hydrogen carbonate presumably existed as an insoluble salt [17]. Only particles entirely composed of C and O were detected in the commercial sterilized substitution fluid and 5% glucose solution. The data shown in Fig. 1 suggest that glucose, which was not completely dissolved, was collected by the membrane filter and measured by SEM–EDS. Therefore, particles in the commercial sterilized substitution fluid and 5% glucose solution were considered to be derived from glucose. Because the average diameter of human pulmonary capillaries is 5 µm [18], particles greater than 5 µm in diameter can obstruct blood flow in the pulmonary capillaries and have a risk of causing complications such as pulmonary embolism [8]. However, glucose is soluble and is not able to obstruct blood flow in the pulmonary capillaries.

Si-containing particles accounted for 83% of 30 particles detected in the final ETRF used for 3 months (the hollow fibers and the housing) and the online prepared substitution fluid sampled from the final ETRF used for 3 months. Si was not detected in particles in the commercial sterilized substitution fluid and 5% glucose solution. Both the commercial sterile acidic and bicarbonate substitution fluids contained SiO2 at 0.1 mg/L. The SiO2 in the 5% glucose solution was below the lower limit of quantification. That information on Si and SiO2 is not included in the commercial sterilized substitution fluid and 5% glucose solution package inserts or Medicine Interview Forms. Therefore, we presume that insoluble particulates in online prepared substitution fluid were Ca and Mg precipitated from dialysate and Si precipitated from dialysis water.

The concordant fluctuation in the SiO2 level of dialysis water and tap water samples shown in Table 4 indicates that variation in the SiO2 in tap water is a factor affecting the SiO2 level in dialysis water. Figure 3 shows the large variability in the SiO2 in tap water depending on the geographic regions. In Japan, sources of tap water include river water and ground water [19]. Water in regions with many volcanoes, such as Hokkaido and Kanto regions, have been known to contain high levels of SiO2 [20]. Moreover, river water in the Chubu region tends to contain low levels of SiO2, that in the Chugoku region contains low levels of SiO2, and that in the Kyushu region contains the highest levels of SiO2 [21]. The SiO2 in tap water sampled in the city of Osaka was the lowest; this is presumably attributable to sediment removal and spring diatom blooms in Lake Biwa, the source of the Yodo River flowing through Osaka City [19, 22]. As shown in Table 4, ETRFs cannot remove SiO2 from the dialysis water and the water treatment equipment removed 89–93% of SiO2 from tap water with a mean removal rate of 91%. Table 5 shows SiO2 in dialysis water estimated with an assumed SiO2 removal rate of 91% from tap water in various geographic regions of Japan. The lowest SiO2 level in dialysis water with an assumed removal rate of 91% from tap water in various geographical regions of Japan was 0.2 mg/L, which is twice as high as the SiO2 in the commercial sterilized substitution fluid. With online HDF necessitates the same (or even higher) quality as that of injectable agents manufactured by pharmaceutical companies [23] because the amount of infusion is much greater in online HDF than in a normal injection. If the target SiO2 in dialysis water is to be set ≤ 0.1 mg/L, the water treatment equipment can be used to reduce SiO2, and additional effective removal of SiO2 can be achieved by combining the water treatment equipment with an electrodeionization (EDI) pure water production system [24].

Table 5 SiO2 in dialysis water estimated with an assumed removal efficiency of 91%

This study is not enough since the relation between patients’ health and Si-containing particles needs to be deeply considered by in vivo study such as using patients’ samples. Si ingested by eating and drinking is excreted in urine, but in patients with renal failure, Si is not excreted in urine and is thought to accumulate [25, 26]. Therefore, serum Si concentrations in hemodialysis patients (64 ± 12 μg/dl) are higher than those in healthy subjects (< 20 μg/dl) [27]. In particular, serum Si concentrations of hemodialysis patients with bone diseases (76 ± 18 μg/dl) and hemodialysis patients with neurological disorders (86 ± 15 μg/dl) are significantly higher than those of hemodialysis patients without complications (64 ± 12 μg/dl), suggesting that high serum Si concentrations in hemodialysis patients are one of the causes of bone diseases and neurological disorders [27]. Atomic spectrometry methods used in the analysis of plasma Si concentrations evaluate the identification and concentration of the element [28]. This analytical method can evaluate the identification and concentration of Si, but it cannot distinguish the form of Si. Multiple forms of Si in water need to be considered, including soluble silica, ionic silica, soluble silicates, colloidal silica, silicates, inorganic and organic substances containing Si, and biological materials containing Si [29,30,31]. Removal of these Si-containing substances also requires consideration of the method of removal depending on the form of Si, such as removal with ion exchange resins if the Si is in ionic form, or removal with membrane devices such as reverse osmosis membranes if the Si is in colloidal form [32]. In addition, when silica is taken into the body other than by ingestion through food or drink (e.g., by inhalation or intravenous injection), it invariably shows some toxicity [33]. Based on these literature findings, the effects on hemodialysis patients regarding Si should be considered comprehensively, including serum Si concentration, the form of Si taken into the hemodialysis patient’s body, and the uptake pathway of silica taken into the hemodialysis patient’s body. Based on the results of the current study, magnesium silicates (particles consisting of Mg and Si) are present in the online prepared substitution fluid and may serve as a trigger of the thrombus formation [34]. SiO2 in the commercial sterilized substitution fluid and 5% glucose solution are ≤ 0.1 mg/L, and Ca, Mg, and Si were not detected in particles in commercial sterile substitution fluid and 5% glucose solution. Therefore, even if two ETRFs are connected in series, additional caution should be exercised for the online prepared substitution fluid because it contains insoluble particulate matters of CaCO3, amorphous SiO2, and magnesium silicates.

Conclusion

We evaluated particles in online prepared substitution fluid obtained from a peracetic acid-cleaned CDDS and particles attached to the final ETRF (hollow fibers and the housing) using SEM/EDS. We also measured SiO2 contained in tap water collected from various regions of Japan, the commercial sterilized substitution fluid, and 5% glucose solution. The measurement results with particles in the online prepared substitution fluid and particles attached to the final ETRF revealed that Si-containing particles accounted for 83% of 30 particles detected in the online prepared substitution fluid and the final ETRF. The SiO2 in tap water for producing dialysis water varied among the geographical regions from which the samples were collected. The commercial sterile substitution fluids contained SiO2 at 0.1 mg/L. The SiO2 in the 5% glucose solution was below the lower limit of quantification.

That information on Si and SiO2 is not included in the powdery dialysate concentrates package inserts or Medicine Interview Forms. Therefore, we presume that insoluble particulate matters in online prepared substitution fluid were Ca and Mg precipitated from dialysate and Si precipitated from dialysis water. Moreover, variation in the SiO2 in tap water is a factor affecting the SiO2 level in dialysis water. If the formation of silicates in the online prepared substitution fluid is considered and the target SiO2 in dialysis water is to be set ≤ 0.1 mg/L, SiO2 in tap water available in hospitals and dialysis centers should be known before specifications of instruments for dialysis fluid preparation are determined.

Availability of data and materials

The datasets used and/or analyzed in the current study are available from the corresponding author on reasonable request.

Abbreviations

A:

Acid

B:

Bicarbonate

JP:

Japanese Pharmacopoeia

JSDT:

Japanese Society for Dialysis Therapy

Si:

Silicon

HDF:

Hemodiafiltration

ETRF:

Endotoxin-retentive filter

CDDS:

Central dialysis fluid delivery system

SEM:

Scanning electron microscopy

EDS:

Energy-dispersive X-ray spectrometry

SiO2 :

Silica

RO:

Reverse osmosis

ICP-OES:

Inductively coupled plasma optical emission spectroscopy

O:

Oxygen

C:

Carbon

Ca:

Calcium

Mg:

Magnesium

SIRS:

Systemic inflammatory response syndrome

EDI:

Electrodeionization

References

  1. Sakuma K, Uchiumi N, Sato S, Aida N, Ishimatsu T, Igoshi T, et al. Experience of using heat citric acid disinfection method in central dialysis fluid delivery system. J Artif Organs. 2010;3:145–50.

    Article  Google Scholar 

  2. The Japanese Pharmacopoeia. 17th ed. Japan: Ministry of Health, Labour and Welfare; 2016.

  3. Mineshima M, Kawanishi H, Ase T, Kawasaki T, Tomo T, Nakamoto H, et al. 2016 update Japanese Society for Dialysis Therapy Standard of fluids for hemodialysis and related therapy. Ren Replace Ther. 2018;4:15.

    Article  Google Scholar 

  4. Wada S, Maruyama T, Wakikawa T, Yamada A, Sasaki S. Laser particle monitor for dialysis fluid. Jpn J Clin Dial (Rinsho Toseki). 2002;18:1333–7 (in Japanese).

    Google Scholar 

  5. Masakane I, Taniguchi M, Nakai S, Tsuchida K, Wada A, Ogata S, et al. Annual dialysis data report 2016, JSDT renal data registry. Ren Replace Ther. 2018;4:45.

    Article  Google Scholar 

  6. Ledebo I, Nystrand R. Defining the microbiological quality of dialysis fluid. Artif Organs. 1999;23:37–43.

    Article  CAS  Google Scholar 

  7. Kawanishi H, Moriishi M, Takahashi N, Tsuchiya S. The central dialysis fluid delivery system (CDDS): is it specialty in Japan? Ren Replace Ther. 2016;2:1.

    Article  Google Scholar 

  8. American Society for Parenteral and Enteral Nutrition. Safe Pract Parent Nutr JPEN. 2004;28:S39-70.

    Google Scholar 

  9. Matsumoto T, Okabayash K, Yamakawa T. Powdered dialysates and silicon (Si). Jpn J Clin Dial (Rinsho Toseki). 2004;20:1195–7 (in Japanese).

    Google Scholar 

  10. Pharmaceutical and Medical Device Regulatory Science Society of Japan: The Japanese Pharmacopoeia Technical Information 2016. Tokyo: JIHO; 2016:243–259 (in Japanese).

  11. Japan Water Works Association: Standard Methods for the Examination of Water II -3–28. Tokyo: Kogusuri Printing Co. Ltd; 2011:210–215 (in Japanese).

  12. Sasse M, Dziuba F, Jack T, Köditz H, Kaussen T, Bertram H, et al. In-line filtration decreases systemic inflammatory response syndrome, renal and hematologic dysfunction in pediatric cardiac intensive care patients. Pediatr Cardiol. 2015;36:1270–8.

    Article  Google Scholar 

  13. Masakane I, Kawanishi H, Mineshima M, Takemoto Y, Uchino J, Hoshino T, et al. Japanese Society for Dialysis Therapy: 2011 JSDT standard on the management of endotoxin retentive filter for dialysis and related therapies. Ther Apher Dial. 2013;17:229–40.

    Article  Google Scholar 

  14. Japan Society of Corrosion Engineering: Corrosion Handbook. Tokyo: Maruzen Publishing; 2000: 159–160 (in Japanese).

  15. Ito K, Sekido K, Kawade M. Deposition properties of magnesium silicate scale in cooling water systems. J Water Waste (Yosui to Haisui). 2016;58:677–84 (in Japanese).

    CAS  Google Scholar 

  16. Kodama T. Water quality index for the evaluation of Ccrrosivity to facilities. Jpn J Water Pollut Res. 1987;10:97–101 (in Japanese).

    Article  Google Scholar 

  17. Guyton AC, Hall JE. Text of medical physiology. 10th ed. Philadelphia: W.B. Sauders Company; 2000. p. 456–7.

    Google Scholar 

  18. Kuno T. AFBF. Jpn J Clin Dial (Rinsho Toseki). 2017; 33:540-541 (in Japanese).

  19. Morii F, Kobayashi J, Miyake Y. Chemical studies on the natural waters. Nogaku kenkyu. 1982;59:171–88 (in Japanese).

    CAS  Google Scholar 

  20. Japan Society of Corrosion Engineering: Corrosion Handbook. Tokyo: Maruzen Publishing;2000:648 (in Japanese).

  21. Kobayashi J. A chemical study on the average quality and characteristics of river waters of Japan. Nogaku Kenkyu. 1961;48:63–103 (in Japanese).

    Google Scholar 

  22. Nishida S, Inosaka N, Tsugita K, Nakatani Y. Dissolved silicate transport in artificial system in lower yodo river basin. J Jpn Soc Civ Eng Ser B1 Hydr Eng. 2015;71:1207–12 (in Japanese).

    Google Scholar 

  23. Minakuchi J. Recent progress in artificial kidneys. Shikoku Acta Med. 2017;73:199–206 (in Japanese).

    CAS  Google Scholar 

  24. Ase T, Watabe T, Sato T. Enhanced production of water for haemodialysis using electrodeionization. Sep Sci Technol. 2017;52:332–43.

    Article  CAS  Google Scholar 

  25. Berlyne G, Dudek E, Adler AJ, Seidmon M. Silicon metabolism; The basic facts in renal failure. Kidney Int. 1985;28(Suppl 17):175–7.

    Google Scholar 

  26. Adler AJ, Berlyne MG. Silicon metabolism: Renal handing in chronic renal failure patients. Nephron. 1986;44:36–9.

    Article  CAS  Google Scholar 

  27. Hosokawa S, Oyamaguchi A, Yoshida O. Relationship between trace element abnormalities and chronic hemodialysis. J Jpn Soc Dial Ther. 1990;23:371–5 (in Japanese).

    Article  Google Scholar 

  28. Fifield FW, Kealy D. Principles and practice of analytical chemistry. 5th ed. Oxford: Blackwell Science; 2000. p. 284–353.

    Google Scholar 

  29. Hiroshima City Institute of Public Health: Annual report of Hiroshima City Institute of Public Health. Hiroshima: Hiroshima City; 2000; 28: 94–95 (in Japanese).

  30. Iwasaki I, Katsura T, Tarutani T. So-called silica content of natural waters. Bull Chem Soc Jpn. 1951;24:227–30.

    Article  CAS  Google Scholar 

  31. Agui W, Shimoyamada H, Abe M, Ogino K. Removal of colloidal silica from water with anion exchange resins. J Jpn Water Works Assoc. 1994;4:176–82.

    CAS  Google Scholar 

  32. Ishii N. Optimal purification technology of ultrapure water for instrumental analysis. Bunseki Kagaku (Japan Analyst). 2011;60:103–13 (in Japanese).

    Article  CAS  Google Scholar 

  33. Iler KR. The chemistry of silica: solubility, polymerization, colloid and surface properties, and biochemistry. New York: Wiley; 1979. p. 172–8.

    Google Scholar 

  34. Gatti AM, Montanari S. Retrieval analysis of clinical explanted vena cava filters. J Biomed Mater Res B Appl Biomater. 2006;77:307–14.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

We would like to extend our most sincere appreciation to Eiichi Nishikawa, PhD, for their cooperation in data collection.

Funding

No funding was received.

Author information

Authors and Affiliations

Authors

Contributions

MI and KT designed the study. MI and MY contributed to data collection. MI performed the analysis. MI and AY wrote the original draft. All authors reviewed and approved the final draft of the manuscript.

Corresponding author

Correspondence to Motoyuki Imai.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Imai, M., Yamamoto, M., Togo, K. et al. Particulates formed in dialysis fluid: effect of silica in tap water. Ren Replace Ther 8, 44 (2022). https://doi.org/10.1186/s41100-022-00433-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41100-022-00433-4

Keywords