Table 4 Comparison of removal amounts for different solutes and albumin leakage (n = 20)

From: Efficacy of polymethylmethacrylate membrane hemodiafilter Filtryzer® PMF™-21A in improving pruritus in hemodialysis patients: a prospective interventional study

 

FIX-210E eco

PMF-21A

p value

FIX-210S eco

PMF-21A

p value

(n = 9)

(n = 11)

BUN, mg

11,783.3 ± 1438.9

10,483.3 ± 2216.4

0.198

13,813.6 ± 2667.6

13,540.9 ± 2213.0

0.744

Cr, mg

1275.0 ± 404.4

1318.3 ± 370.0

0.482

1659.5 ± 464.9

1682.7 ± 317.2

0.733

IP, mg

623.3 ± 146.0

643.3 ± 205.1

0.593

867.3 ± 158.3

842.7 ± 146.8

0.584

β2-MG, mg

190.9 ± 43.5

3.9 ± 1.7

< 0.001

198.6 ± 51.8

3.6 ± 1.6

< 0.001

α1-MG, mg

75.8 ± 18.2

6.0 ± 13.3

< 0.001

154.4 ± 47.7

1.2 ± 3.9

< 0.001

Alb, g

2.3 ± 0.5

1.6 ± 0.6

< 0.01

5.7 ± 1.0

1.5 ± 0.4

< 0.001

  1. Data (mg or g) are presented as mean ± standard deviation. P values were calculated using paired t-test
  2. α1-MG α1-microglobrin, Alb albumin, β2-MG β2-microglobrin, BUN blood urea nitrogen, Cr creatinine, IP inorganic phosphorus, PMF-21A Filtryzer® PMF™-A