Renal Replacement Therapy

Open Access

Associations between fluid removal and number of B-lines, peak early mitral inflow wave velocity, and inferior vena cava dimensions in hemodialysis patients

  • Kazuo Kimura1, 2,
  • Katsuya Kajimoto3,
  • Shigeru Otsubo1, 4Email author,
  • Takashi Akiba2 and
  • Kosaku Nitta1
Renal Replacement Therapy20173:54

https://doi.org/10.1186/s41100-017-0141-8

Received: 28 August 2017

Accepted: 27 November 2017

Published: 6 December 2017

Abstract

Background

The inferior vena cava (IVC) dimensions represent the right ventricular preload, whereas the peak early mitral inflow wave velocity (peak E-velocity) represents the left ventricular preload. On the other hand, B-lines represent extravascular lung water. The aim of this study was to evaluate possible acute changes in the IVC dimensions, peak E-velocity, and number of B-lines during hemodialysis therapy.

Methods

A total of 55 consecutive patients receiving maintenance hemodialysis were enrolled in this study. We performed echo-graphic examinations at three time points (just after the start, during the middle, and just before the end of the hemodialysis therapy). We then investigated the changes in the IVC dimensions, peak E-velocity, and number of B-lines.

Results

The peak E-velocity decreased from 80 ± 26 cm/s at the start of the therapy to 58 ± 22 cm/s during the middle and 51 ± 21 cm/s at the end of the therapy. The IVC dimensions also decreased from 15 ± 4 mm at the start of the therapy to 12 ± 3 mm during the middle and 11 ± 3 mm at the end of the therapy. The number of B-lines also decreased from 12 ± 5 at the start of the therapy to 9 ± 4 during the middle and 5 ± 3 at the end of the therapy. The changes in the peak E-velocity and IVC dimensions were significantly greater during the first half of the dialysis period than during the second half of the dialysis period (P < 0.0001 and P < 0.0001, respectively). On the other hand, the changes in the number of B-lines during these periods were significantly smaller during the first half of the dialysis period than during the second half of the dialysis period (P = 0.0016).

Conclusions

We showed that the peak E-velocity and the IVC dimensions were reduced mainly during the first half of the dialysis period, while the number of B-lines showed a significant decrease mainly during the last half of the dialysis period. Even if the IVC dimensions are reduced sufficiently, caution is needed as lung congestion may still exist.

Keywords

B-linesEarly diastolic filling velocityEarly diastolic filling waveHemodialysisInferior vena cavaLung ultrasound

Background

Chronic fluid overloading frequently occurs in hemodialysis patients, so volume assessment during hemodialysis is a primary, and often challenging, goal of nephrologists. Volume overload is directly associated with hypertension, increased arterial stiffness, left ventricular hypertrophy, heart failure, and ultimately higher rates of mortality and morbidity [1].

Lung ultrasound has recently been shown to be a useful, noninvasive technique for the assessment of extravascular lung water [2]. The most commonly observed finding was a comet tail artifact fanning out from the lung-wall interface and spreading upwards to the edge of the screen, previously named a “B-line” [3]. In patients with heart failure, the number of B-lines was correlated with the degree of extravascular lung water [4, 5]. Lung ultrasound is also reportedly useful for the accurate evaluation of dry weight and fluid status in hemodialysis patients [2, 6, 7].

The inferior vena cava (IVC) dimensions represent the right ventricular preload [8], whereas the peak early mitral inflow wave velocity (peak E-velocity) represents the left ventricular preload [9, 10]. On the other hand, B-lines represent extravascular lung water. The aim of this study was to evaluate possible acute changes in IVC dimensions, peak E-velocity, and B-lines during hemodialysis therapy.

Methods

A total of 55 patients receiving maintenance hemodialysis at Sekikawa Hospital and in whom more than 1 kg was removed during hemodialysis were enrolled in this study. Patients with lung disease or chest injury were excluded from this study.

Clinical data including age, sex, duration of hemodialysis therapy, presence of diabetes mellitus and/or hypertension and/or dyslipidemia complications, and the results of biological examinations were collected from the patients’ clinical records. Hypertension was defined as a systolic blood pressure of 140 mmHg or higher, a diastolic blood pressure of 90 mmHg or higher, and/or the current use of antihypertensive drugs. Diabetes mellitus was defined as a fasting glucose level ≥ 126 mg/dL, a nonfasting glucose level ≥ 200 mg/dL, or the use of medication. Dyslipidemia was defined as a low-density lipoprotein cholesterol level ≥ 140 mg/dL, a high-density lipoprotein cholesterol level < 40 mg/dL, a triglyceride level ≥ 150 mg/dL, or the use of medication.

A peripheral blood sample was obtained before hemodialysis during the first session of the week. The serum N-terminal pro-brain natriuretic peptide (NT-proBNP) level in the pre-dialysis blood sample was measured using an electrochemiluminescence immunoassay on an Elecsys platform (Roche, Basel, Switzerland).

A lung ultrasound examination was performed during the first session of the week (at the same time as the peripheral blood sample preparation) using Vscan® (GE Healthcare, Japan), which is a hand-held ultrasound device with a wide-bandwidth phased-array probe (1.7–3.5 MHz) [11, 12]. Bilateral scanning of the anterior and lateral chest walls was performed with the patient in a supine position. An intercostal scan with a maximum extension of the visual pleural line was performed. The chest wall was divided into eight areas (two anterior and two lateral areas per side), and one scan was obtained for each area [13]. The anterior zone of the chest wall was designated from the sternum to the anterior axillary line and was then divided into upper and lower halves (from the clavicle to the third intercostal spaces and from the third space to the diaphragm). The lateral zone was positioned from the anterior axillary line to the posterior axillary line and was also divided into the upper and lower halves. The investigators attempted to detect comet tail artifacts fanning out from the lung-wall interface and spreading to the edge of the screen, which were previously named as B-lines (Fig. 1) [5, 6]. The total number of B-lines was estimated. Echocardiographic measurements were obtained at the same time, and the IVC dimensions and peak E-velocity were estimated. We performed echo-graphic examinations three times (just after the start, just in the middle (for example, 2 h after the start of hemodialysis therapy if the treatment time is 4 h), and just before the end of the hemodialysis therapy). We then investigated the changes in the IVC dimensions, peak E-velocity, and B-lines. This study was conducted in accordance with the principles of the Declaration of Helsinki and was permitted by the research ethics committee of Sekikawa Hospital (Approved No. H2705). The data were expressed as the means ± SD or median (interquartile range, IQR). Paired t tests were used to examine changes in the variables.
Fig. 1

B-lines—comet tail artifacts fanning out from the lung-wall interface and spreading upwards to the edge of the screen

All the statistical calculations were performed using JMP 5.1 software. P values less than 0.05 were considered statistically significant.

Results

The patient background characteristics are shown in Table 1. The mean age was 75 ± 11 years. Elderly patients were common in this study. Diabetic nephropathy was the major cause of end-stage kidney disease. Hypertension, diabetes mellitus, and dyslipidemia were present in 89.1, 47.3, and 16.4% of the study participants, respectively. The results of the biochemistry analyses, including the NT-proBNP level, are shown in Table 2. The serum albumin level was relatively low (3.1 ± 0.5 g/dL), and the serum NT-proBNP level was relatively high (13,650 [4901–35,825] pg/mL) among the study participants. Table 3 shows the background characteristics of the echocardiographic indices. The mean ejection fraction was 60% ± 12%.
Table 1

Background characteristics of the study participants

Characteristic

Quantity

Gender (M/F)

26/29

Age (year)

75 ± 11

Duration of HD (year)

6.5 ± 8.8

Primary cause of ESKD, n (%)

 Chronic glomerulonephritis

8 (14.5)

 Diabetic nephropathy

26 (47.3)

 Nephrosclerosis

13 (23.6)

 Unknown and others

8 (14.5)

Hypertention, n (%)

49 (89.1)

Diabetes mellitus, n (%)

26 (47.3)

Dyslipidemia, n (%)

9 (16.4)

Atrial fibrillation, n (%)

8 (14.5)

History of PCI or CABG

8 (14.5)

History of valve replacement

2 (3.6)a

Body weight (kg)

52.0 ± 10.9

Systolic blood pressure (mmHg)

147 ± 25

Diastolic blood pressure (mmHg)

74 ± 15

Heart rate (beat/min)

75 ± 13

Mean ± SD, median (interquartile range)

HD hemodialysis, ESKD end-stage kidney disease, PCI percutaneous coronary intervention, CABG coronary artery bypass graft

aBoth patients underwent aortic valve replacement therapy

Table 2

Background characteristics of laboratory data

Characteristic

Quantity

Albumin (g/dL)

3.1 ± 0.5

Urea nitrogen (mg/dL)

50.2 ± 18.3

Creatinine (mg/dL)

7.36 ± 2.56

Sodium (mEq/L)

137 ± 4

C-reactive protein (mg/dL)

0.27 (0.10–0.86)

Hemoglobin (g/dL)

10.4 ± 1.2

NT-proBNP (pg/mL)

13,650 (4901–35,825)

Mean ± SD, median (interquartile range)

NT-proBNP N-terminal pro-brain natriuretic peptide

Table 3

Background characteristics of the echocardiographic indices

Parameter

Value

LVDd (mm)

46 ± 6

LVDs (mm)

31 ± 7

EF (%)

60 ± 14

Max WT (mm)

12 ± 2

Mean ± SD

LVDd left ventricular diastolic diameter, LVDs left ventricular systolic diameter, EF ejection fraction, WT wall thickness

The treatment time was 3.7 ± 0.4 h, and 2.1 ± 0.9 kg water was removed. The blood pressure and pulse rate were 146 ± 25/74 ± 15 mmHg and 75 ± 13 beats/min at the first echo examination, 134 ± 24/72 ± 14 mmHg and 72 ± 16 beats/min at the second, 132 ± 27/73 ± 16 mmHg and 73 ± 15 beats/min at the final examination. Table 4 shows the changes in the echocardiographic indices. The peak E-velocity decreased from 80 ± 26 cm/s at the start of the therapy to 58 ± 22 cm/s during the middle and 51 ± 21 cm/s at the end of the therapy. The IVC dimensions also decreased from 15 ± 4 mm at the start of the therapy to 12 ± 3 mm during the middle and 11 ± 3 mm at the end of the therapy. The number of B-lines also decreased from 12 ± 5 at the start of the therapy to 9 ± 4 during the middle and 5 ± 3 at the end of the therapy (also shown in Fig. 2).
Table 4

Changes in echocardiographic indices

Parameter

Just after the start (1st)

Middle (2nd)

Just before the end (3rd)

P (1st vs 2nd)

P (2nd vs 3rd)

E (cm/s)

80 ± 26

58 ± 22

51 ± 21

< 0.0001

< 0.0001

A (cm/s)a

87 ± 30

77 ± 22

72 ± 20

0.0006

0.0039

E/A

0.95 ± 0.39

0.72 ± 0.26

0.68 ± 0.26

< 0.0001

0.0083

Dct (msec)

255 ± 97

271 ± 92

281 ± 94

Ns

Ns

Ea (cm/s)

5.3 ± 1.7

5.4 ± 1.6

5.6 ± 1.6

0.0459

Ns

TRPG (mmHg)

4.5 ± 12.4

2.8 ± 9.5

2.4 ± 9.0

0.0237

Ns

IVC-ex (mm)

15 ± 4

12 ± 3

11 ± 3

< 0.0001

< 0.0001

IVC-in (mm)

11 ± 5

5 ± 3

3 ± 3

< 0.0001

< 0.0001

IVC collapsibility (%)

31 ± 19

61 ± 18

74 ± 20

< 0.0001

< 0.0001

B-lines

12 ± 5

9 ± 4

5 ± 3

< 0.0001

< 0.0001

Mean ± SD

E peak early diastolic filling velocity, A peak Atrial filling velocity, Dct deceleration time, Ea lateral early diastolic mitral annular velocities, TRPG tricuspid regurgitation pressure gradient, IVC inferior vena cava dimensions, ex exhalation, in inspiration

aExclude eight atrial fibrillation

Fig. 2

Changes in echocardiographic indices. a Changes of peak E-velocity. b Changes of IVC dimensions. c Changes of B-lines. Peak E-velocity peak early mitral inflow wave velocity, IVC inferior vena cava

Table 5 shows a comparison of the changes in values between the middle to pre-dialysis periods and the post-dialysis to middle periods. The changes in the peak E-velocity and the IVC dimensions were significantly greater during the first half of the dialysis period than during the second half of the dialysis period (P < 0.0001 and P < 0.0001, respectively). On the other hand, the changes in the B-lines between these periods was significantly smaller during the first half of the dialysis period than during the second half of the dialysis period (P = 0.0016).
Table 5

Comparison of the changes in values between the middle to pre-dialysis periods and the post-dialysis to middle periods

 

2nd–1st

3rd–2nd

P value

ΔE (cm/s)

− 23 ± 14

− 7 ± 6

< 0.0001

ΔIVC-ex (mm)

− 4 ± 3

− 1 ± 2

< 0.0001

ΔB-lines

− 3 ± 3

− 4 ± 2

0.0016

Mean ± SD

1st just after the start, 2nd middle, 3rd just before the end, E peak early diastolic filling velocity, IVC inferior vena cava dimensions, ex exhalation

Figure 3 shows a representative case in this study. The peak E-velocity and IVC dimensions decreased mainly during the first half of the dialysis period, while the number of B-lines decreased steadily throughout the dialysis period.
Fig. 3

Case of a 71 year-old female, chronic kidney disease 5D (due to diabetic nephritis). Change of peak E-velocity, IVC dimensions, and B-lines were shown. B-mode images of actual E and A velocity were also shown. Remarkable decrease of peak E-velocity was observed compared with peak atrial filling velocity. Peak E-velocity peak early mitral inflow wave velocity, IVC inferior vena cava

Discussion

We found that the peak E-velocity and IVC dimensions decreased significantly mainly during the first half of the dialysis period, while the number of B-lines decreased significantly mainly during the last half of the dialysis period.

Lung ultrasound is a novel, well-validated technique that allows reliable estimates of lung water in clinical practice [14]. B-lines can be evaluated anywhere (including extreme environmental conditions using pocket-sized instruments to detect high-altitude pulmonary edema), anytime (during dialysis), by anyone (even a novice sonographer after 1 h of training), and on anybody (since the chest acoustic window usually remains patent even when an echocardiography is not feasible) [15].

The peak E-velocity represents the flow to the left ventricle, which occurs after the left ventricle diastole, reflecting the preload status [9, 10]. Ultrafiltration during hemodialysis results in a marked reduction in the circulating blood volume with a concomitant reduction in the left atrial pressure. Acharya et al. examined the parameters of left ventricular diastolic function among hemodialysis patients [16]. A comparison of the pre- and post-hemodialysis peak mitral E and A velocities showed a decrease in E-velocity, whereas the change in A velocity was not significant. Ultrafiltration during hemodialysis causes a rapid reduction in the preload, resulting in decreased early left ventricular diastolic filling without a change in the atrial phase of filling, thereby reducing the calculated E/A ratio [16]. We confirmed the change in the peak E-velocity and revealed a greater reduction during the first half of the dialysis period than during the second half of the dialysis period. We also confirmed that E/A ratio was declined during the hemodialysis therapy.

The IVC dimensions are strongly associated with the right atrial pressure and blood volume and therefore reflect the intravascular volume accurately. However, a major limitation of IVC dimension measurements is that ultrasound might not be sensitive enough to detect rapid volume decreases in HD patients because it reflects the intravascular filling grade [17, 18]. Indeed, similar to the change in the peak E-velocity, we showed that the IVC dimensions decreased mainly during the first half of the dialysis period, while the B-lines, which represent lung congestion, decreased significantly mainly during the last half of the dialysis period. This finding may be largely dependent on the time lag in plasma refilling. Alexiadis et al. reported that the number of B-lines has better discriminative power for predicting over- and underhydration, as determined by the IVC dimensions [2]. Mallamaci et al. suggested that the detection of extracellular water in the lung compartment constitutes a reliable tool for the prognosis of over-hydration and is thus amenable to intervention [19]. So, we should be careful that even if the IVC dimensions are sufficiently reduced, lung congestion may still exist and should be treated.

Our study had some limitations. The sample size was relatively small, and the study was performed at a single institution. Activities of daily living of some patients were poor, so we could not examine height of patients. In addition, the limitations of lung ultrasound are essentially patient-dependent. Obese patients may be more difficult to examine because of the thickness of their ribcages and soft tissues. Furthermore, in this analysis, we performed ultrasound examinations only at the end of the hemodialysis therapy. However, it has been suggested that post-dialysis assessments should ideally occur 2 h after hemodialysis because of the re-equilibration of interstitial and intravascular compartments [20, 21]. Accordingly, further study is needed to evaluate the relationship between fluid removal and the timing of post-dialysis assessments in hemodialysis patients. Factors which could affect the change of the number of B-lines are important for the clinician. The change of B-lines was negatively related to the number of B-lines at the start of the hemodialysis therapy (r 2 = 0.609, P < 0.0001, data not shown). We should examine the association between the change of the number of B-lines and various factor under the same number of B-lines at the start of hemodialysis therapy, but we could not because of the small number of study participant. Even though ultrasound of extravascular lung water is reported as a new standard for pulmonary congestion [15], there is still no gold standard of examination to evaluate excess body water. In this study, we could not show other possible parameters such as chest X-ray or bioimpedance spectroscopy. Accordingly, further study will be needed to clarify this finding. Another limitation of this study is that we could not show the optimal number of B-lines after hemodialysis therapy. B-lines within physiological range may be present. We have previously reported that B-lines are associated with low serum level of albumin and low body weight [6]. So, optimal number of B-lines may be different among hemodialysis patients.

Conclusions

We showed that the peak E-velocity and IVC dimensions were reduced mainly during the first half of the dialysis period, while the number of B-lines decreased significantly mainly during the last half of the dialysis period. Even if the IVC dimensions are sufficiently reduced, caution is required as lung congestion may still exist.

Abbreviations

E-velocity: 

Peak early mitral inflow wave velocity

IVC: 

Inferior vena cava peak

NT-proBNP: 

N-terminal pro-brain natriuretic peptide

Declarations

Acknowledgements

The authors are very grateful to the dialysis staff who understood the clinical importance of this study and who provided high-quality data in Sekikawa Hospital.

Funding

This study was not supported by any grants or funding.

Availability of data and materials

The datasets during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Authors’ contributions

KK planned the study, searched the literature, assessed the studies, extracted the data, analyzed the data, and prepared the article. KK and SO searched the literature, assessed the studies, and assisted in the article preparation. KK performed the lung echo. KN and TA assisted in the article preparation. All authors read and approved the final manuscript.

Ethics approval and consent to participate

This study was approved by the research ethics committee of Sekikawa Hospital (Approved No. H2705) and was conducted under the Declaration of Helsinki.

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Authors’ Affiliations

(1)
Department of Medicine, Kidney Center, Tokyo Women’s Medical University, Tokyo, Japan
(2)
Department of Nephrology, Sekikawa Hospital, Tokyo, Japan
(3)
Department of Cardiology, Sekikawa Hospital, Tokyo, Japan
(4)
Department of Blood Purification, Tohto Sangenjaya Clinic, Tokyo, Japan

References

  1. Tonelli M, Wiebe N, Culleton B, et al. Chronic kidney disease and mortality risk: a systematic review. J Am Soc Nephrol. 2006;17:2034–47.View ArticlePubMedGoogle Scholar
  2. Alexiadis G, Panagoutsos S, Roumeliotis S, Stibiris I, Markos A, Kantartzi K, et al. Comparison of multiple fluid status assessment methods in patients on chronic hemodialysis. Int Urol Nephrol. 2017;49:525–32.View ArticlePubMedGoogle Scholar
  3. Lichtenstein D. Pneumothorax and introduction to ultrasound signs in the lung. In: Heilmann U, Wilbertz H, Gosling A, editors. General ultrasound in the critically ill. 1st ed. Heidelberg: Springer-Verlag; 2005. p. 105–15.Google Scholar
  4. Agricola E, Bove T, Oppizzi M, Marino G, Zangrillo A, Margonato A, et al. “Ultrasound comet-tail images”: a marker of pulmonary edema: a comparative study with wedge pressure and extravascular lung water. Chest. 2005;127:1690–5.View ArticlePubMedGoogle Scholar
  5. Kajimoto K, Madeen K, Nakayama T, Tsudo H, Kuroda T, Abe T. Rapid evaluation by lung-cardiac-inferior vena cava (LCI) integrated ultrasound for differentiating heart failure from pulmonary disease as the cause of acute dyspnea in the emergency setting. Cardiovasc Ultrasound. 2012;10:49–51.View ArticlePubMedPubMed CentralGoogle Scholar
  6. Kuzuhara S, Otsubo S, Kajimoto K, Akiba T, Nitta K. Association between B-lines detected during lung ultrasound and various factors in hemodialysis patients. Ren Replace Ther. 2017;3:17.View ArticleGoogle Scholar
  7. Noble VE, Murray AF, Capp R, Sylvia-Reardon MH, Steele DJ, Liteplo A. Ultrasound assessment for extravascular lung water in patients undergoing hemodialysis: time course for resolution. Chest. 2009;135:1433–9.View ArticlePubMedGoogle Scholar
  8. Simonson JS, Schiller NB. Sonospirometry: a new method for noninvasive estimation of mean right atrial pressure based on two-dimensional echographic measurements of the inferior vena cava during measured inspiration. J Am Coll Cardiol. 1988;11:557–64.View ArticlePubMedGoogle Scholar
  9. Triulzi MO, Castini D, Ornaghi M, Vitolo E. Effects of preload reduction on mitral flow velocity pattern in normal subjects. Am J Cardiol. 1990;66:995–1001.View ArticlePubMedGoogle Scholar
  10. Sztajzel J, Ruedin P, Monin C, Stoermann C, Leski M, Rutishauser W, et al. Effect of altered loading conditions during haemodialysis on left ventricular filling pattern. Eur Heart J. 1993;14:655–61.View ArticlePubMedGoogle Scholar
  11. Cardim N, Fernandez Golfin C, Ferreira D, Aubele A, Toste J, Cobos MA, et al. Usefulness of a new miniaturized echocardiographic system in outpatient cardiology consultations as an extension of physical examination. J Am Soc Echocardiogr. 2011;24:117–24.View ArticlePubMedGoogle Scholar
  12. Liebo MJ, Israel RL, Lillie EO, Smith MR, Rubenson DS, Topol EJ. Is pocket mobile echocardiography the next-generation stethoscope? A cross-sectional comparison of rapidly acquired images with standard transthoracic echocardiography. Ann Intern Med. 2011;155:33–8.View ArticlePubMedPubMed CentralGoogle Scholar
  13. Volpicelli G, Elbarbary M, Blaivas M, Lichtenstein DA, Mathis G, Kirkpatrick AW, et al. International liaison committee on lung ultrasound (ILC-LUS) for international consensus conference on lung ultrasound (ICC-LUS). International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012;38:577–91.View ArticlePubMedGoogle Scholar
  14. Zoccali C. Lung ultrasound in the management of fluid volume in dialysis patients: potential usefulness. Semin Dial. 2017;30:6–9.View ArticlePubMedGoogle Scholar
  15. Picano E, Pellikka PA. Ultrasound of extravascular lung water: a new standard for pulmonary congestion. Eur Heart J. 2016;37:2097–104.View ArticlePubMedPubMed CentralGoogle Scholar
  16. Acharya P, Ranabhat K, Trikhatri Y, Manandhar DN, Sharma SK, Karki P. Effect of preload reduction by haemodialysis on doppler indices of diastolic function in patients with end-stage renal disease. Kathmandu Univ Med J (KUMJ). 2008;6:98–101.Google Scholar
  17. Feissel M, Michard F, Faller JP, Teboul JL. The respiratory variation in inferior vena cava diameter as a guide to fluid therapy. Intensive Care Med. 2004;30:1834–7.View ArticlePubMedGoogle Scholar
  18. Brennan JM, Ronan A, Goonewardena S, Blair JE, Hammes M, Shah D, et al. Handcarried ultrasound measurement of the inferior vena cava for assessment of intravascular volume status in the outpatient hemodialysis clinic. Clin J Am Soc Nephrol. 2006;1:749–53.View ArticlePubMedGoogle Scholar
  19. Mallamaci F, Benedetto FA, Tripepi R, Rastelli S, Castellino P, Tripepi G, et al. Detection of pulmonary congestion by chest ultrasound in dialysis patients. JACC Cardiovasc Imaging. 2010;3:586–94.View ArticlePubMedGoogle Scholar
  20. Katzarski KS, Nisell J, Randmaa I, Danielsson A, Freyschuss U, Bergstrom J. A critical evaluation of ultrasound measurement of inferior vena cava diameter in assessing dry weight in normotensive and hypertensive hemodialysis patients. Am J Kidney Dis. 1997;30:459–65.View ArticlePubMedGoogle Scholar
  21. Kouw PM, Kooman JP, Cheriex EC, Olthof CG, de Vries PM, Leunissen KM. Assessment of postdialysis dry weight: a comparison of techniques. J Am Soc Nephrol. 1993;4:98–104.PubMedGoogle Scholar

Copyright

© The Author(s) 2017

Advertisement